Calibrating Monitor Spectrophotometers: Steps For Accurate Color Readings

how to properly calibrate a monitor spectrophotometer

Calibration is essential to ensure the accuracy of data collected by a monitor spectrophotometer. The process involves a series of steps, including warming up the device, selecting the wavelength for calibration, zeroing the machine, and performing a photometric accuracy scan. This scan can be done manually with the spectrophotometer controls or more efficiently with accompanying software if available. During calibration, it is crucial to handle the equipment with care, wearing powder-free gloves to prevent damage to filters. Additionally, the use of a reference solution and a Fiber Optic Cable for further calibration is also recommended.

shundigital

Warm up the spectrophotometer for 45 minutes

Warming up a spectrophotometer is an essential step in the calibration process. It is important to allow the spectrophotometer to warm up for 45 minutes to ensure that it is functioning optimally and that accurate readings can be obtained. During the warm-up period, the machine stabilises and reaches a consistent temperature, which is necessary for precise measurements.

While the spectrophotometer is warming up, you can prepare your samples and the workspace. Put on the required powder-free gloves to prevent damage to the filters and handle the calibration standards with care. Refer to the owner's manual to identify the specific wavelengths that need to be tested for your particular spectrophotometer. This information is crucial as it ensures that the machine is calibrated across its entire working range.

The warm-up period is also an ideal time to clean the cuvettes or test tubes. Use deionized water to rinse each cuvette thoroughly. If you are using quartz cuvettes for UV-visible spectrophotometry, handle them with extra care as they can be quite expensive. Avoid touching the sides of the cuvette that the light will pass through. If you accidentally touch these sides, use a kimwipe to wipe down the cuvette, as these are designed to prevent scratching.

Additionally, you can use this time to load your samples into the cuvettes. Ensure that the laser-producing light passes through the liquid and not an empty part of the container for accurate readings. If you are using a pipette, use a new tip for each sample to prevent cross-contamination.

By efficiently utilising the 45-minute warm-up period, you can ensure that your spectrophotometer is ready for accurate calibration and that your samples and workspace are prepared according to the standard protocols.

shundigital

Calibrate dark

Calibrating a monitor spectrophotometer in a dark environment is essential for accurate measurements and reliable results. Here are the steps to properly "Calibrate Dark":

Firstly, ensure you have the necessary equipment: the spectrophotometer, its manual, calibration standards and their certificate, and powder-free gloves. The gloves are crucial to protect the filters from damage during handling. Before beginning, refer to the manual to understand the specific wavelengths your spectrophotometer will test, as these may vary between devices. Some operate only in the VIS range, for example.

Now, let's get started with the procedure:

  • Warm up the spectrophotometer for approximately 45 minutes. This is an important step to ensure the device is ready for accurate calibration.
  • Select the wavelength you want to calibrate, such as 465 nm.
  • Determine if you need a dedicated blank for the machine based on the type of filter you're using. If your standard came with a blank, insert it now. If not, leave the cuvette holder empty and close the lid.
  • Zero the spectrophotometer so that the display shows all zeros.
  • Remove the blank if you used one.
  • Insert the NIST calibration standard and close the cover.
  • Record the reading and compare it to the data on the certificate of calibration.

If the data matches, congratulations! Your spectrophotometer is now calibrated for dark measurements. However, if the data doesn't match, there are a few troubleshooting steps you can take:

  • Check the uncertainty of the calibration standard on the Certificate of Calibration.
  • If the machine is still out of spec, refer to the owner's manual for photometric accuracy tolerance information. If this isn't available, contact the manufacturer for assistance.
  • Add the tolerance value to the filters to obtain the extended tolerance.
  • If the numbers still don't match, the issue may lie with the spectrophotometer or the calibration standards. Try using the standard on another spectrophotometer to isolate the problem.
  • If the standard doesn't work on a different machine, the NIST calibration filter may require cleaning or re-calibration.

By following these steps, you can ensure that your monitor spectrophotometer is properly calibrated for dark conditions, enabling you to obtain precise and reliable measurements.

shundigital

Prepare a reference solution

Before calibrating the spectrometer, prepare a reference solution by filling a cuvette 3/4 with distilled water. Place the cuvette containing the reference solution into the spectrometer so that a clear side is facing the white light source icon. Only handle the cuvette on the lined sides and wipe the smooth sides clean with a lint-free cloth.

Cover the sample well with your finger to block any ambient light from entering the detector. A checkmark will appear when the calibration is complete.

shundigital

Calibrate reference

Calibration is essential to ensure the accuracy of your spectrophotometer. Here is a detailed guide on how to properly calibrate the device, with a specific focus on the "Calibrate Reference" step:

Before beginning the calibration process, gather the necessary tools and materials. These include the spectrophotometer itself, the manual for the device, NIST spectrophotometer calibration standards, a certificate of calibration, and powder-free gloves to prevent damage to the filters during handling.

Now, let's delve into the steps for calibration:

This step is part of the two-point calibration process, which involves referencing one point to the absence of light (dark) and the other point to the transmitted light through the solute used in your solutions. Firstly, prepare a reference solution by filling a cuvette 3/4 with distilled water. This cuvette will be used for calibration.

Next, connect the spectrometer to the Spectrometry software. Go to the "Analyze Solution" page and click "Calibrate Dark" from the bottom of the screen. This will turn off the light sources in the spectrometer to perform the initial calibration for the absence of light. Ensure that you cover the sample well during this step to block any ambient light from entering the detector. A check mark will appear once the calibration is complete.

Now, place the cuvette containing the reference solution into the spectrometer, ensuring that a clear side faces the white light source icon. It is important to only handle the cuvette on the lined sides, and wipe the smooth sides clean with a lint-free cloth to avoid any interference with the calibration process.

Finally, click "Calibrate Reference" from the bottom of the screen. This will initiate the calibration process for the transmitted light reference point. Once the calibration is complete, a check mark will appear, indicating that your spectrophotometer has been successfully calibrated for these two reference points.

By following these steps, you can ensure that your spectrophotometer is accurately calibrated and ready for use. This process improves the accuracy of the data collected by your device, making it a crucial step before any measurements or experiments.

shundigital

Calibrate wavelength

Calibration is necessary to correct for systematic error. This is done by comparing experimentally determined wavelengths to wavelengths obtained from the literature. To calibrate a spectrophotometer, you will need the manual for the spectrophotometer, calibration standards, a certificate of calibration, and powder-free gloves.

Firstly, warm up the spectrophotometer for 45 minutes. Then, select the wavelength to calibrate, for example, 465 nm. If your standard came with a blank, insert it now, if not, leave the cuvette holder empty and close the lid. Zero the spectrophotometer so that the display shows all zeros. Remove the blank (if used), insert the calibration standard, and close the cover. Record the reading and compare it to the data on the certificate of calibration.

To calibrate the wavelength of your spectrometer, you will need a light source that produces spectral lines, an optical fibre, and a calculator or spreadsheet program that performs second-order linear regressions. After placing OOIBase into Scope Mode, take a spectrum of your light source. Adjust the A/D conversion frequency or integration time until there are several peaks on the screen that are not off-scale. Move the cursor to one of the peaks and carefully position it at the point of maximum intensity. Record the pixel number that is displayed in the status bar. Repeat this step for all of the peaks in your spectrum.

Using your spreadsheet, create a table with the exact wavelength of the spectral lines in the first column. In the second column, place the observed pixel number. In the third column, place the pixel number squared. In your spreadsheet program, find the functions to perform linear regressions. Select the true wavelength as the dependent variable (Y) and select both the pixel number and pixel number squared as the independent variables (X). Record the Intercept, as well as the First and Second Coefficients.

To calibrate a spectrometer with a CCD array detector, it is necessary to acquire many partial spectra, each at a different angular position of the grating, and splice the partial spectra into a single extended spectrum. This procedure exposes instrument-dependent artifacts. By taking a spectrum of a reference irradiance source and making spectral corrections, these artifacts can be removed.

The calibration of a spectroscope can be done by using a helium discharge tube. Look through the eyepiece of your spectroscope and find the six helium lines. Note the position of these lines with regard to the scale and record their colours and their wavelengths in Angstroms. Readings are dependent on the position of your eye, so one person should take the readings and be consistent with the position of their eye relative to the eyepiece. Prepare a calibration plot by graphing the scale reading obtained experimentally on the x-axis versus the literature wavelength value on the y-axis.

Frequently asked questions

To calibrate a monitor spectrophotometer, you will need the manual for the spectrophotometer, NIST spectrophotometer calibration standards, a certificate of calibration, powder-free gloves, and the spectrophotometer itself.

The first step is to warm up the spectrophotometer for 45 minutes.

If the data doesn't match, check the uncertainty of the calibration standard on the Certificate of Calibration. If the machine is still out of spec, refer to the owner's manual for the photometric accuracy tolerance. If the numbers still don't match, the problem may lie with the spectrophotometer or the calibration standards.

Written by
Reviewed by
Share this post
Print
Did this article help you?

Leave a comment